Thermo Fisher Scientific
SAM12 Small Articles Monitor
Monitor for contamination checks of small articles and tools, with 4 or 6 detectors and one or two doors.
The SAM12 is available with 4 or 6 detectors and one or two doors, and as an electronics upgrade for older SAM11 monitors. Typical applications include clean tool shops, containment exits, active barriers at the RCA boundary and free release surveys during dismantling.
- Measures fixed, smearable, internal and external gamma contamination simultaneously
- Natural Background Reduction (NBR) discriminates between NORM and man-made radiation
- Cobalt Coincidence Monitoring (CCM), insensitive to fluctuating gamma background
- No cooling fan and completely self contained: no peripherals needed to set up or configure
- Integral 12 V power pack gives 8 hours of operation if AC supplies are lost
No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

Authorised distributor in SwedenAll products from Thermo Fisher Scientific →
Technical data
| Detectors | 4 or 6 BC-412 plastic scintillation detectors, 1451 cm² each |
|---|---|
| Radiation type | Gamma |
| Energy range | 50 keV to 2 MeV |
| MDA (10 s, 5 cm lead, 0.1 µSv/h background) | 6 detectors: Co-60 56 Bq, Cs-137 120 Bq; 4 detectors: Co-60 70 Bq, Cs-137 180 Bq |
| Typical 4π efficiency in centre of chamber | 6 detectors: Co-60 57 %, Cs-137 28 %; 4 detectors: Co-60 40 %, Cs-137 16 % |
| Article monitoring time | 3 to 300 s |
| Measuring volume (H × W × D) | 381 × 381 × 457 mm |
| Dimensions (H × W × D) | 1114 × 687 × 837 mm (922 mm deep for 6 detector, 2 door option) |
| Lead shielding | 2.5 or 5 cm |
| Weights | 670 kg nett (1" lead) or 1380 kg nett (2" lead) |
| Power | 85 to 264 VAC, 47 to 63 Hz; integral 12 V power pack for 8 hours |
| Environment | 0 °C to +45 °C, up to 95 % RH non-condensing |
Source: Thermo Fisher Scientific, SAM12 datasheet (2019). Full specifications are available in the documents.