LAM12 Large Article Monitor
Thermo Fisher Scientific

LAM12 Large Article Monitor

Monitor for contamination checks and clearance of large articles, with four detectors around a 750 × 600 × 600 mm measuring chamber.

The LAM12 helps ensure that large articles leaving the controlled area of a nuclear facility are within regulatory limits for radioactive contamination. It shares its electronics and touchscreen with the SAM12 and is available in four versions: standard, with a low energy copper and aluminium filter set, with a weigher scale, or with both. Older LAM4 monitors can be upgraded with LAM12 electronics.

  • Measures fixed, smearable, internal and external gamma contamination simultaneously
  • Measures down to clearance levels below 0.4 Bq/g for gamma-emitting radionuclides
  • Natural Background Reduction (NBR) discriminates between NORM and man-made radiation
  • Cobalt Coincidence Monitoring (CCM), insensitive to fluctuating gamma background
  • QuickScan reduces counting time when articles are clearly clean or clearly contaminated
  • Optional weigher scale, so the specific activity (Bq/g) can be assessed and displayed

No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

Authorised distributor in SwedenAll products from Thermo Fisher Scientific →

Technical data

DetectorsFour BC-412 plastic scintillation detectors, 600 × 300 × 50 mm each (front door, back and two sides)
Radiation typeGamma
Energy range50 keV to 2 MeV
Typical 4π efficiency in centre of chamber⁶⁰Co 24 %, ¹³⁷Cs 12 %; low energy option ²⁴¹Am 2.5 %; CCM ⁶⁰Co 0.2 %
MDA (30 s, 0.1 µSv/h background)⁶⁰Co 95 Bq, ¹³⁷Cs 185 Bq; low energy option ²⁴¹Am 1600 Bq
MDA (30 s, 1 µSv/h background)⁶⁰Co 300 Bq, ¹³⁷Cs 650 Bq
Minimum specific activity detectable< 0.1 Bq/g of ⁶⁰Co averaged over 5 kg
Spatial uniformity of response±30 % at 68 % confidence, for ¹³⁷Cs
Measuring volume (H × W × D)750 × 600 × 600 mm
Dimensions (H × W × D)1400 × 935 × 975 mm
Lead shielding25 mm, 4π
Weight1350 kg net
Weigher option0.5 to 60 kg, 570 × 570 mm platform, 0.02 kg resolution
Power85 to 264 VAC, 47 to 63 Hz, 65 VA; integral 12 V power pack for 8 hours
Digital I/O connectionsEthernet and 4 USB
Operational temperature0 °C to +45 °C; up to 95 % RH non-condensing, IP50

Source: Thermo Fisher Scientific, LAM12 datasheet (2019). Full specifications are available in the documents.