FHT 192 ionization chamber with styrofoam protector
Thermo Fisher Scientific

FHT 192 Ionization Chamber

Wide range pressurised ionization chamber for high precision ambient equivalent dose rate measurements, for the FHT 6020, FH 40 G and SVG2.

The FHT 192 connects to the FHT 6020 area monitor or the FH 40 G survey meter, for measurements at high energy accelerators or X-ray sources.

  • 30 keV to 7 MeV energy range and 100 nSv/h to 1 Sv/h dose rate range
  • Based on the PTB-approved medium-pressure chamber FHT 191 N, used as a reference standard detector by many organisations
  • Chamber geometry with excellent angular and energy dependence (¾ sphere angular response)
  • Styrofoam protection minimises currents induced by temperature changes
  • Integrated preamplifier fits the FHT 6020, FH 40 G and SVG2 – high dose option with remote preamplifier

No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

Authorised distributor in SwedenAll products from Thermo Fisher Scientific →

Technical data

Detector typePressurised ionization chamber
Measured variableAmbient equivalent dose rate H*(10)
Measuring range100 nSv/h – 1 Sv/h
Energy range30 keV – 7 MeV (± 30 %)
Gamma responsiveness250 fA/µSvh⁻¹
Angular acceptance−45° < Φ < +90° (270° solid angle)
OverloadUp to 50 Sv/h overload indication
Operating gas (permanently filled)Noble gas – nitrogen – mixture, 7000 hPa (absolute value)
Volume4.25 litres
Length500 mm incl. polystyrene cap (detector part 270 mm)
Diameter208 mm incl. polystyrene cap
Weightapprox. 3.25 kg (7.1 lb)

Source: Thermo Fisher Scientific, FHT 192 product specifications (2007) and product page 425354050. Full specifications are available in the documents.