Thermoluminescent dosimetry (TLD): readers, dosimeter cards and TLD materials for passive dosimetry in radiation protection monitoring.
The Harshaw readers are used worldwide in IEC, NVLAP, DOELAP and ANSI accredited facilities. They are complemented by multi-element TLD cards with Teflon capsulated TL material mounted in an aluminium card, tamper-resistant holders with optimised filtration for energy discrimination, and sealed disposable or sterilisable extremity dosimeters such as finger rings and wrist bands.
Harshaw TLD materials are available un-mounted (chips, disks, rods, micro-cubes and powders) and mounted in multi-element cards with two to four elements. The cards are worn in coloured holders with filtration that discriminates between deep dose, shallow dose and dose to the lens of the eye, and every card is uniquely barcoded. For calibration there is the Harshaw Model 2210 table top irradiator, which moves samples into and out of the beam on a rotating disk, and a ⁹⁰Sr internal irradiator built into the automatic readers as an alternative to a large ¹³⁷Cs source. For higher volumes there are also the Harshaw 5500 for un-mounted elements, the Harshaw 6600 Plus for up to 200 cards per run and the Harshaw 8800 Plus for up to 1400 cards per run, and for extremity monitoring the EXT-RAD and DXT-RAD dosimeters.
No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

| Range | Seven decades (both readers) |
|---|---|
| Photronics linearity | Less than 1 % deviation |
| Repeatability | Less than 2 % variation (1 standard deviation of 10 sequential measurements at 1 mGy ¹³⁷Cs) |
| Warm-up time | 30 minutes (3500), 20 minutes (6600 Plus) |
| Readout time | 35 sec/dosimeter (3500); TLD cards 70 per hour, EXT-RAD Chipstrates 140 per hour, DXT-RAD Ringlets 280 per hour (6600 Plus) |
| Dosimeter heating | Planchet heating to 400 °C, or 600 °C with the high temperature model (3500); non-contact nitrogen gas heating (6600 Plus) |
| Card identification | Codabar, Interlaced 2 of 5, Code 39, Code 128 alphanumeric (6600 Plus) |
| Electrical | 110 or 120 VAC ±10 %, 60 Hz; 220 or 240 VAC ±10 %, 50 Hz |
| Operating temperature range | 15 °C to 40 °C |
| Weight | 25 kg (3500), 70 kg (6600 Plus) |
| Height × width × depth (3500) | 31 × 32 × 47 cm |
| TLD-100/-600/-700 (LiF:Mg,Ti) | Linear range 10 µGy to 1 Gy; fading < 20 % in 3 months, < 5 % in 3 months corrected |
| TLD-100H/-600H/-700H (LiF:Mg,Cu,P) | Linear range 1 µGy to 10 Gy; fading negligible in 3 months, < 5 % up to 2 years |
| TLD-200 (CaF₂:Dy) | Linear range 0.1 µGy to 10 Gy; fading 10 % in 1st 24 hr, 16 % total in 2 weeks |
| TLD-400 (CaF₂:Mn) | Linear range 0.1 µGy to 100 Gy; fading 8 % in 1st 24 hr, 12 % total in 2 weeks |
| Available forms | Chips, MicroCube, square rod, rod, unsorted chips, powder, pelletised chip, pelletised disk (varies by material) |
| Multi-element TLD cards | 2 to 4 Teflon capsulated elements in an aluminium card; Hp(10), Hp(0.07) and Hp(3); personnel (holders 8806, 8814, 8825, 8840, 8850, 9900) and environmental (8855) dosimeters |
| Model 2210 chip irradiator | ⁹⁰Sr/⁹⁰Y, 18 MBq (0.5 mCi); rotating sample disk; 155 × 260 × 265 mm (H × W × D), 9.5 kg |
| ⁹⁰Sr internal irradiator | Approx. 0.5 mCi, nominal exposure rate 10 mR/s; for 6600 Plus and 8800 Plus |
Source: Thermo Fisher Scientific, Harshaw 3500, Harshaw 6600 Plus and Harshaw TLD Materials and Dosimeters brochures (2025) and product pages HARSHAWTLD2210 and 21989. Full specifications are available in the documents.