Baltic Scientific Instruments

HPGe Spectrometer with Lead Shield

Complete laboratory spectrometer with a coaxial HPGe detector in a lead shield, for radionuclide analysis of environmental and food samples.

The spectrometer is commonly used to measure gamma radiation in rocks, minerals, sludge, slag, soil, plants, sediments and food. The integrated lead shielding minimises background radiation, improving detection sensitivity and accuracy for laboratory analysis.

  • Radionuclide identification and determination of their specific activity
  • Separate and simultaneous measurement of activity of 100 radionuclides
  • Detection limit 0.5 Bq/kg for Cs-137 with 1 hour measurement time
  • Low instrumental background, in standard spectrometric, low-background or ultra low-background grades
  • Lead thicknesses 100 mm and 150 mm

No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

Technical data

Detection limit for Cs-137, 1 hour0.5 Bq/kg
Cs-137 measurement error, 1 hour20 %
Absolute sensitivity (30 % detector)4.5 × 10⁻³ pulse/quantum
Instrumental background, 40 keV to 3 MeV5 × 10⁻⁴ pulse/(keV × s)
Lead thickness100 mm or 150 mm
CryostatVertical or U-type
CoolingLN2, Monolith or Nicole
Complete setHPGe detection unit, lead shield with support, MCA, GammaPRO, cable set and documentation