AirTrack aerosol monitoring station in a cabinet
Baltic Scientific Instruments

AirTrack Aerosol Monitoring Station

Automatic monitoring station for continuous measurement of alpha, beta and gamma radionuclides in airborne aerosols, with automatic filter changer.

AirTrack is designed for autonomous operation and delivers real-time data for industrial, research and emergency applications. Gamma detectors are fully interchangeable without design modifications, and all operations can be controlled from a remote computer. The gamma channel is automatically stabilised using the K-40 peak.

  • Continuous automatic measurement of alpha, beta and gamma radionuclide concentrations in airborne aerosols
  • Silicon detector for alpha/beta; SrI₂, LaBr₃, CeBr₃, NaI or electrically cooled HPGe detector for gamma
  • Removable cartridge with 48 (60) filter capacity, up to 2 months of autonomous operation
  • Compensation for radon progeny and gamma background, and evaluation according to ISO 11929
  • Data transfer via LAN, USB and 4G in the ANSI N42.42/EURDEP format

No prices online – you will receive a quotation with price and delivery time. Need accessories? Mention them in your request.

Technical data

Detector typeAlpha/beta: 900 mm² silicon ion-implanted detector. Gamma: SrI₂(Eu), optionally LaBr₃(Ce), CeBr₃, NaI(Tl) or HPGe with electrical cooling
Energy measurement rangeAlpha 2 MeV to 10 MeV; beta 0.05 MeV to 2 MeV; gamma 0.05 MeV to 3 MeV (SrI₂) or 3 keV to 3 MeV (HPGe)
Concentration measurement rangeAlpha 10⁻² to 10⁵ Bq/m³; beta 10⁻¹ to 10⁵ Bq/m³; gamma 10⁻² to 10⁶ Bq/m³
Energy resolutionAirTrack-Sr: < 3.2 % at 661.7 keV (SrI₂(Eu) Ø 38.1 × 38.1 mm); AirTrack-Ge: < 1 keV at 122 keV, < 2 keV at 1.33 MeV
MCA1024 channels (alpha), 4096 channels (gamma)
FiltersAny Ø 47 mm disk filter; automatic replacement when activity or differential pressure thresholds are exceeded
Update period1 minute
Power supply230 V, 50 Hz
Operation temperature+5 °C to +40 °C

Source: Baltic Scientific Instruments, AirTrack brochure (bsi.lv, retrieved November 2026). Full specifications are available in the documents.